4-14 | Thermox CEM O2/TM
attempt a standard calibration again. If you get an excess cal error during
primary calibration, then the furnace temperature set point cannot be in-
creased any further to prevent damage to the cell housing and/or furnace
assembly.
Failure limits for standard calibration -
+/- 10 mVs of the anticipated mVs for the gas concentration applied (air 0
mV expected).
+/- 5 mVs of the anticipated mVs for the gas concentration applied (2%
O2 applied, 44 mV expected at 615°C cell temp) as referenced to the actual
span gas mV reading.
T/C ratio adjustment limits for Primary Calibration -
Cell Temp °C
Max T/C Ratio
Min T/C Ratio
615
1.3
0.7
650
1.24
0.74
695
1.22
0.777
824
1.139
0.861
790
same as 824°C
For a cell operating temperature of 615°C, the T/C ratio low limit is 0.7
and the high limit is 1.3, which corresponds to a thermocouple indicated
temperature range of 473°C to 878°C.
If the analyzer is configured for a cell operating temperature of 615°C, the
highest possible operating temperature for the analyzer at after a primary
calibration is 878°C.
This menu option allows you to perform a primary calibration on the
zirconium oxide oxygen sensor. Perform a primary calibration on your
system rather than a regular calibration (from the Calibrate Key) in the
following cases:
When you replace your zirconium oxide oxygen cell.
When your system fails a regular calibration.
Содержание Thermox CEM O2/TM
Страница 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Страница 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Страница 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Страница 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Страница 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Страница 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Страница 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Страница 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Страница 137: ...Service and Parts 7 3 This page intentionally left blank...
Страница 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Страница 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Страница 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Страница 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Страница 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...