Installation
|
3-3
Figure 3-1.
Sensor mounting.
Mounting the Sensor
The CEM O2/TM sensor enclosure is a NEMA 12 rated polyurethane
coated carbon steel cabinet that you mount to a wall. Figure 3-1 provides
wall mounting dimensions for the sensor. Provide at least six inches clear-
ance above the sensor so you can later replace the cell. The cell is replaced
by lifting it through the top of the sensor enclosure.
The sensor should be located as close to the sampling system as possible.
The ambient temperature must be in the range of -18 to 50°C (0 to 122°F).
The sensor relies on a convection principle to flow sample gas over the
oxygen cell. Therefore, it must be mounted vertically.
Содержание Thermox CEM O2/TM
Страница 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Страница 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Страница 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Страница 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Страница 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Страница 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Страница 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Страница 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Страница 137: ...Service and Parts 7 3 This page intentionally left blank...
Страница 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Страница 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Страница 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Страница 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Страница 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...