2-2 | Thermox CEM O2/TM
Communications
RS-485, 2-way addressable.
Environment
Ambient Temp: 14 to 122° F (-10 to 50° C)
Relative Humidity: 0% to 80%, non-condensing
Max. Altitude: 2000 meters
IEC Installation (Overvoltage) Category: II
IEC Pollution Degree 2
hazardous Area
UL Listed for NEC Class I, Division 2
Enclosure
19” rack mount standard. Optional panel mount, UL Type 4 (NEMA 4
[IP56]) and UL Type 4 (NEMA 4 X [IP56]) enclosures available. All are UL
Listed for NEC Class I, Division 2 areas.
Calibration
Store last calibration and verification data. Selectable calibration gas run
time and process recovery time. Oxygen cell lifetime extender. Single gas
verify that analyzer is within calibration limits. Timed automatic calibra-
tion with optional remote calibration unit.
Power Requirements
Nominal 115-230 VAC ±10%, 47-63 Hz, 75 VA max.
System Compliance
EMC Directive 2004/108/EC
Low Voltage Directive 73-23/EEC
Содержание Thermox CEM O2/TM
Страница 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Страница 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Страница 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Страница 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Страница 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Страница 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Страница 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Страница 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Страница 137: ...Service and Parts 7 3 This page intentionally left blank...
Страница 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Страница 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Страница 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Страница 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Страница 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...