Specifications | 2-3
Sensor
Operating Range
1 PPM O
2
to 100% O
2
Accuracy
Percent: ±0.75% of measured value or ±0.05% O
2
, whichever is greater
PPM: ± 2% of reading or 0.5 PPM O
2
absolute, whichever is greater.
Response Time
< 10 seconds at 1.0 l/min (2 scfh) to 90% of 2-decade step change.
Repeatability
Percent: ± 0.5% of reading or 0.1% O
2
absolute, whichever is greater.
PPM: ± 0.5% of reading or 0.1 PPM O
2
absolute, whichever is greater.
Drift
< 0.1% of cell output per month (< 0.005% O2 with 2% O2 applied)
Max. Inlet Temperature
400° F (204° C)
Sample Flow
1.0 L/min (2 scfh)
Environment
Ambient Temp: 0°F to 122°F (-18°C to 50°C)
Relative Humidity: 0% to 80%, non-condensing
Содержание Thermox CEM O2/TM
Страница 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Страница 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Страница 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Страница 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Страница 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Страница 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Страница 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Страница 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Страница 137: ...Service and Parts 7 3 This page intentionally left blank...
Страница 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Страница 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Страница 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Страница 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Страница 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...