Installation
|
3-13
Wiring
Remove AC mains power from the controller and sensor before per-
forming wiring.
Connections to the control unit are made through the wiring card (see
Figure 3-14). The wiring card is located on the front, bottom of the control
unit for all weatherproof versions and for the general purpose wall mount
version (see the “Control Unit Mounting” Section earlier in this chapter
for help). For the general purpose rack and panel mount versions, the wir-
ing card is located on the rear of the control unit (Figure 3-14 shows the
orientation of the wiring card for general purpose rack and panel mount
versions - for all other versions, the wiring card is 180 degrees from this
view).
To access the wiring card for all general purpose versions of the control
unit, unscrew the four screws on the wiring card chassis cover plate. For
the weatherproof versions of the control unit, first open the front door of
the weatherproof enclosure to gain access to the wiring card shield. Re-
move the three screws on this shield to gain access to the wiring card.
Any screw terminals on the wiring card not described in this section
are reserved for future use.
Figure 3-14.
Wiring Card (rack/panel
mount orientation).
NOTE
Содержание Thermox CEM O2/TM
Страница 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Страница 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Страница 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Страница 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Страница 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Страница 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Страница 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Страница 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Страница 137: ...Service and Parts 7 3 This page intentionally left blank...
Страница 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Страница 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Страница 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Страница 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Страница 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...