4-16 | Thermox CEM O2/TM
Sensor Configuration
This menu option allows you to select either normal operations or hy-
drocarbon upset condition operations. The hydrocarbon upset condition
option allows the system to recognize a hydrocarbon upset condition and
trigger an alarm if that condition is reached.
If you do not want to use the Hydrocarbon Upset option, select Nor-
mal for the Sensor Configuration.
Configurating the Sensor
1. Select Sensor Config from the Setup Key menu.
2. Select Normal or Hydrocarbon Upset.
3. If you select Hydrocarbon Upset in Step 2 above, you are then
prompted to enter a millivolt value that, if exceeded, triggers the hy-
drocarbon upset condition.
If the oxygen reading drops to 0 PPM, and the millivolt value you
entered is reached or exceeded, the system displays a low oxygen/high
hydrocarbons message rather than the oxygen reading (assuming
oxygen is displayed on the control unit). It also trips Alarm 4.
If the oxygen reading is 0, but the cell millivolt threshold value has not
been reached, the system will display a low oxygen/low hydrocarbon
message.
If you select Hydrocarbon Upset in Step 2 above, Alarm 4 becomes
fixed as a high oxygen alarm and also acts as a hydrocarbon upset
alarm. You may need to reenter your alarm setpoints and configura-
tion.
Once a hydrocarbon upset condition is reached (and assuming you select-
ed Hydrocarbon Upset), the display switches from displaying the oxygen
reading to displaying a hydrocarbon condition text message.
NOTE
NOTE
Содержание Thermox CEM O2/TM
Страница 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Страница 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Страница 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Страница 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Страница 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Страница 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Страница 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Страница 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Страница 137: ...Service and Parts 7 3 This page intentionally left blank...
Страница 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Страница 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Страница 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Страница 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Страница 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...