Calibration | 5-13
The verify display then appears:
O2 xx.x Set xx.x
Press ENTER to Verify
5. Press the Enter key once the O
2
reading on the far left of the first
display line stabilizes. The reading on the far right of the first display
line shows the set point value you entered using the Configure Verify
menu option. This set point value should match the value on your
verify gas cylinder.
If your O
2
reading on the far left of the display has stabilized, yet is not
near the set point value on the right of the display, check the cal gas value
you entered and ensure that this value matches the cal gas cylinder for
the gas you applied to the cal gas inlet port of the sensor.
Then, if you specified a recovery time, the system will display the recov-
ery time remaining:
Recovery xx:xx
System Verifying Cal
The system then returns to monitoring process readings.
Содержание Thermox CEM O2/TM
Страница 12: ...xii Thermox CEM O2 TM This page intentionally left blank...
Страница 45: ...Installation 3 23 Figure 3 20 Current output connections example...
Страница 47: ...Installation 3 25 Figure 3 21 Series 2000 standard alarm connections...
Страница 49: ...Installation 3 27 Figure 3 22 Series 2000 controller RCU connections...
Страница 51: ...Installation 3 29 Figure 3 23 Series 2000 2 wire RS 485 communications connections...
Страница 53: ...Installation 3 31 Figure 3 24 Series 2000 4 wire RS 485 communications connections...
Страница 56: ...3 34 Thermox CEM O2 TM This page intentionally left blank...
Страница 116: ...5 26 Thermox CEM O2 TM This page intentionally left blank...
Страница 137: ...Service and Parts 7 3 This page intentionally left blank...
Страница 139: ...Service and Parts 7 5 Figure 7 1 Cell replacement...
Страница 143: ...Service and Parts 7 9 Figure 7 3 Furnace replacement...
Страница 162: ...B 4 Thermox CEM O2 TM This page intentionally left blank...
Страница 166: ...C 4 Thermox CEM O2 TM This page inentionally left blank...
Страница 168: ...D 2 Thermox CEM O2 TM This page intentionally left blank...