A-30 | Thermox Series 2000 WDG-HPII / HPIIC - ATEX
Starting/Monitoring a Calibration: An Example
This section provides an example of how to initiate and then monitor a
calibration.
1. Check that no calibration or verification is currently occurring by check-
ing the Calflag variable (Calflag=3).
2. Send the Calibrate command to the IQ analyzer (G00).
3. If an acknowledge is returned (A<cr>), the calibration command is
accepted. If not accepted, the sensor may not be at a proper operating
temperature or may already be performing a system calibration.
4. Read the Gas_ctr variable to determine the position in the calibration
cycle. Use this variable to determine when to activate solenoids (if not
controlling solenoids using the IQ analyzer).
5. Read the Cal_flag variable to determine when the cal cycle is finished
(Calflag=3).
6. Read the calibration values (for example, cal_span and cal_zero vari-
ables).
Содержание Thermox 2000 ATEX Series
Страница 18: ...1 8 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 24: ...2 6 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 68: ...3 44 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 128: ...5 28 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 149: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor...
Страница 151: ...Service and Parts 7 5 Figure 7 2 Cell replacement...
Страница 153: ...Service and Parts 7 7 Figure 7 3 Furnace thermocouple replacement...
Страница 157: ...Service and Parts 7 11 Figure 7 5 Furnace replacement...
Страница 159: ...Service and Parts 7 13 Figure 7 6 Combustibles detector replacement Hot Wire...
Страница 196: ...B 4 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 202: ...D 2 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...