3-10 | Thermox Series 2000 WDG-HPII / HPIIC - ATEX
Customer flange mounting
The sensor may also be mounted to a 3" 300# customer-supplied flange as
shown in Figure 3-5. In this case, you won’t need to use the Thermox mount-
ing plate with the attached pipe nipple. Then mount the sensor to the
process as described in the “Mounting the Sensor to Process” section be-
low.
Avoid overtightening mounting bolts if using a raised face flange.
This will prevent damaging the backplate of the sensor.
!
CAUTION
Figure 3-5.
Customer flange
mounting.
EXHAUST TUBE
GASKET (BY AMETEK)
SENSOR BOLTED TO MOUNTING FLANGE
LENGTH MINIMUM
NECESSITATED BY
INSTALLATION
REQUIRES 1" (2.54cm) MIN
INSULATION AROUND EXPOSED
PORTION OF NIPPLE
PROCESS WALL
FILTER
(INSIDE EXHAUST TUBE)
SAMPLING PROBE
INSULATING SLEEVE
SIDE VIEW
SINGLE STUD
AT "12 O'CLOCK"
POSITION
PROPER FLANGE ORIENTATION
CUSTOMER'S 3"-300# FLANGE
(REDUCING FLANGE IF REQUIRED)
CONNECTED TO 2 1/2"MIN SCH 40
PIPE ATTACHED TO WALL BY CUSTOMER
Содержание Thermox 2000 ATEX Series
Страница 18: ...1 8 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 24: ...2 6 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 68: ...3 44 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 128: ...5 28 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 149: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor...
Страница 151: ...Service and Parts 7 5 Figure 7 2 Cell replacement...
Страница 153: ...Service and Parts 7 7 Figure 7 3 Furnace thermocouple replacement...
Страница 157: ...Service and Parts 7 11 Figure 7 5 Furnace replacement...
Страница 159: ...Service and Parts 7 13 Figure 7 6 Combustibles detector replacement Hot Wire...
Страница 196: ...B 4 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 202: ...D 2 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...