Calibration | 5-11
You are then prompted to calibrate the zero gases:
O2 xx.x % Set xx.x %
Cmb yyyy Set 0
Press ENTER to Zero
5. Press the Enter key once the zero gas reading(s) on the left of the
display has stabilized. If you don’t have the combustibles option, you
only need to wait until the O
2
zero gas value stabilizes. If you have
the combustibles option, wait until all readings stabilize. The reading
on the far right of the first display line shows the setpoint value you
entered for the O
2
zero gas using the Cal Gas Values menu option.
This O
2
zero gas setpoint value should match the cal gas cylinder for
the O
2
zero gas connected to your RCU.
If you have the combustibles option, you are also prompted to calibrate
the combustibles span gas:
O2 xx.x %
Cmb yyyy Set yyyy
Press ENTER to Span
6. Press the Enter key once the combustibles span gas reading(s) stabi-
lize on the far left of the display.
Then, if you specified a recovery time, the system will display the recov-
ery time remaining:
Recovery xx:xx
System Calibrating
The system then returns to monitoring process readings.
Содержание Thermox 2000 ATEX Series
Страница 18: ...1 8 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 24: ...2 6 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 68: ...3 44 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 128: ...5 28 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 149: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor...
Страница 151: ...Service and Parts 7 5 Figure 7 2 Cell replacement...
Страница 153: ...Service and Parts 7 7 Figure 7 3 Furnace thermocouple replacement...
Страница 157: ...Service and Parts 7 11 Figure 7 5 Furnace replacement...
Страница 159: ...Service and Parts 7 13 Figure 7 6 Combustibles detector replacement Hot Wire...
Страница 196: ...B 4 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 202: ...D 2 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...