Serial Communications
|
A-25
62/98
Cal_variable
The calculated percent value that 0 mv
would correspond to with the cell that
is currently being used. This value
can vary between different cells. It
is calculated during calibration and
used in the oxygen calculations.
Read only
63/99 Option_flags
Option flags (1=true, 0=false)
bit 0
– combustible detector installed
bit 1
– Data module installed
IQ
Dual current output
installed
S2
bit 2
– 2 span value RCU
IQ
auto cal option installed
S2
bit 3
– Methane detector installed
IQ
3-in-1 option installed
S2
bit 4
– Dig – ma module installed
IQ
bit 5
– HART module installed
IQ
bit 6
– RS485 module installed
IQ
bit 7
– 3 span value RCU
IQ
Invalid option module
S2
Read only
64/100
Tspan_mv
Cell mv value measured with the span
gas flowing during last calibration.
Read only
65/101
Tzero_mv
Cell mv value measured with the zero
gas flowing during calibration.
R/W
66/102 Mthvalue
Read
only
67/103 M_store
Read
only
68/104
Cal_temp
Last calib temperature
R/W
69/105
CJTemp
Cold Junction temperature
Read only
6A/106
Box_temp
Box Temperature
Read only
6B/107
Cmb_Kfact
Combustible K Factor
R/W
6C/108
Mth_Kfact
Methane K Factor
IQ
R/W
6D/109
Location
(Hex/Decimal) Variable Name
Description
Access
Содержание Thermox 2000 ATEX Series
Страница 18: ...1 8 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 24: ...2 6 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 68: ...3 44 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 128: ...5 28 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 149: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor...
Страница 151: ...Service and Parts 7 5 Figure 7 2 Cell replacement...
Страница 153: ...Service and Parts 7 7 Figure 7 3 Furnace thermocouple replacement...
Страница 157: ...Service and Parts 7 11 Figure 7 5 Furnace replacement...
Страница 159: ...Service and Parts 7 13 Figure 7 6 Combustibles detector replacement Hot Wire...
Страница 196: ...B 4 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 202: ...D 2 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...