Serial Communications
|
A-1
Serial communications for the Series 2000 (ver 3.3 and later)
and the IQ based software
IQ
applies to IQ sensor based products only
S2
applies to Series 2000 based products only
Communication Protocols
Communication with all analyzers through the RS-485 connection uses
the following word format:
• 8 data bits
• 1 stop bit
• No Parity
In addition, the baud rate is fixed at 9600bps.
The communication protocol is a basic master/slave, command/response
protocol. The master initiates each communication transfer. There is one
master on the communication line, but there can be multiple sensors act-
ing as slaves. Up to 32 sensors can be connected on one network.
Command Message Format
The command message format is as follows:
Where:
Start Character
Node Address
Command Letter
Data
Checksum
End Character
“>” (ASCII
character 3E Hex)
2 hex ASCII
characters (00 to
FF)
(See Command
Section for list of
commands)
(n characaters
- depends on
command)
2 hex ASCII
characters (sum of
all preceding
characters except
the “>” character)
Carriage return
(ASCII character
0D Hex)
SERIAL COMMUNICATIONS
Содержание Thermox 2000 ATEX Series
Страница 18: ...1 8 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 24: ...2 6 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 68: ...3 44 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 128: ...5 28 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 149: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor...
Страница 151: ...Service and Parts 7 5 Figure 7 2 Cell replacement...
Страница 153: ...Service and Parts 7 7 Figure 7 3 Furnace thermocouple replacement...
Страница 157: ...Service and Parts 7 11 Figure 7 5 Furnace replacement...
Страница 159: ...Service and Parts 7 13 Figure 7 6 Combustibles detector replacement Hot Wire...
Страница 196: ...B 4 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 202: ...D 2 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...