3-12 | Thermox Series 2000 WDG-HPII / HPIIC - ATEX
Calibration
Required Calibration Gases and Tubing
See “Specifications” for required pressure/flow of calibration gases.
The span gas is the high gas. The zero gas is the low gas.
Use dry calibration gases that are free of oil and dirt. Stainless steel tubing
is recommended.
The span gas must be 10 times greater than the zero gas.
y
For example, if the zero gas is 1% O
2
, the span gas must be 10% O
2
or
higher.
O
2
span gas- air or from 1.0% to 100% O
2
O
2
zero gas - from .1 to 10% O
2
, balance N
2
Combustibles span gas - 60 to 80% (ppm range) or 40 to 60% (percent
range) of the full combustibles operating range in certified mixtures of
equal parts CO + H
2
, with an equal amount of O
2
or more, balance N
2
.
Example
: If using an operating range of 0 to 2000 ppm combus-
tibles, use 800 ppm CO, 800 ppm H
2
, 1 to 2% O
2
, balance N
2
. If
using a 0 to 1% combustibles range, use 0.3% CO, 0.3% H
2
, 1 to
2% O
2
, balance N
2
.
Connections
Connect instrument air to the sensor purge inlet. Maintain the pressure
and flow as directed on the purge warning label.
The purge apparatus provided by Thermox meets requirements for
hazardous locations under certain conditions.
•
Do not exceed the maximum temperature rating listed in the
Specifications Chapter.
NOTE
!
WARNING
Содержание Thermox 2000 ATEX Series
Страница 18: ...1 8 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 24: ...2 6 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 68: ...3 44 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 128: ...5 28 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 149: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor...
Страница 151: ...Service and Parts 7 5 Figure 7 2 Cell replacement...
Страница 153: ...Service and Parts 7 7 Figure 7 3 Furnace thermocouple replacement...
Страница 157: ...Service and Parts 7 11 Figure 7 5 Furnace replacement...
Страница 159: ...Service and Parts 7 13 Figure 7 6 Combustibles detector replacement Hot Wire...
Страница 196: ...B 4 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 202: ...D 2 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...