3-16 | Thermox Series 2000 WDG-HPII / HPIIC - ATEX
Manual Calibration Connections
If you have a Remote Calibration Unit (RCU), skip the remainder of
this section and proceed to the “RCU Mechanical Installation” sec-
tion that follows.
If you don’t have an RCU, simply connect calibration gases to the calibra-
tion gas inlet port on the sensor (see Figure 3-7). For this manual calibra-
tion setup, you need to plug the calibration inlet port when not being used
for calibrations.
Remote Calibration Unit Mechanical Installation
The ambient temperature range for the RCU is -20 to 70
o
C. Mount the
RCU as close to the sensor as possible (Figure 3-8 shows mounting
dimentions). Shorter calibration plumbing improves response times, re-
duces calibration gas expense, and reduces the chance of contaminants in
the calibration gas plumbing.
For all RCU connections, use 1/4” O.D. tubing. Be sure to use
tubing that is free of oil and direct; we recommend stainless steel
tubing.
Don’t open the RCU enclosure until you verify that the area has been
classified as nonhazardous.
This section is divided into instructions for O
2
-only RCU connections and
O
2
and combustibles RCU connections. Use the appropriate instructions
when making the connections.
NOTE
!
WARNING
NOTE
Содержание Thermox 2000 ATEX Series
Страница 18: ...1 8 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 24: ...2 6 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 68: ...3 44 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 128: ...5 28 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 149: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor...
Страница 151: ...Service and Parts 7 5 Figure 7 2 Cell replacement...
Страница 153: ...Service and Parts 7 7 Figure 7 3 Furnace thermocouple replacement...
Страница 157: ...Service and Parts 7 11 Figure 7 5 Furnace replacement...
Страница 159: ...Service and Parts 7 13 Figure 7 6 Combustibles detector replacement Hot Wire...
Страница 196: ...B 4 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 202: ...D 2 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...