6-2 | Thermox Series 2000 WDG-HPII / HPIIC - ATEX
System Messages
Auto Cal Pending
This message indicates that an automatic calibration is pending. Once the
current calibration or verification is completed, the automatic calibration
will begin. This message will only appear if the Auto Calibration option is
installed and enabled.
Auto Verify Pending
This message indicates that an automatic verification is pending. Once the
current calibration or verification is completed, the automatic verification
will begin. This message will only appear if the Auto Calibration option is
installed and enabled.
Calibration Aborted
This message indicates that the calibration has been aborted by the user, or
the calibration failed because the span gas or zero gas was out of range.
Calibration Required
Once you complete a primary calibration, this message is displayed to in-
form you that a regular calibration is required (after a primary calibration
is performed, you must always then perform a regular calibration). This
message will not clear from the display by removing and restoring power
to the control unit. If you see this message, select the
Calibrate
key and
perform a calibration.
Cal Verify Aborted
This message notifies you that the verification was successfully aborted by
the system as you requested.
Power Down Detected
The system detected that power was removed. For example, a power out-
age would cause this message to display. You can acknowledge this system
message by setting or reading any user programmable value, and this will
clear the message from the display. This message will not clear by remov-
ing and restoring power to the unit.
Содержание Thermox 2000 ATEX Series
Страница 18: ...1 8 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 24: ...2 6 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 68: ...3 44 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 128: ...5 28 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 149: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor...
Страница 151: ...Service and Parts 7 5 Figure 7 2 Cell replacement...
Страница 153: ...Service and Parts 7 7 Figure 7 3 Furnace thermocouple replacement...
Страница 157: ...Service and Parts 7 11 Figure 7 5 Furnace replacement...
Страница 159: ...Service and Parts 7 13 Figure 7 6 Combustibles detector replacement Hot Wire...
Страница 196: ...B 4 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 202: ...D 2 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...