Current Outputs: Other Applications | C-1
CURRENT OUTPUTS: OTHER APPLICATIONS
This appendix describes how to do the following:
•
Modulate external power loop using current outputs.
•
Power current outputs from an external power supply.
Powering Current Outputs from an External Power Supply
The Series 2000 control unit wiring card contains a set of terminals that
can be used to connect an external voltage source to power the standard
current outputs:
VEXT1, 2 +
VEXT1, 2 -
Jumper JP2 on the Series 2000 control unit display module is used to allow
you to define which power source to use. If JP2 is in (factory default), the
control unit uses the internal supply to power the current output circuits.
If you try to use an external voltage at this time, the system will use which-
ever voltage is higher - control unit voltage is normally about 28 volts. If
the JP2 jumper is removed, the external voltage source must be used to
power the current output circuits. The external voltage source must be no
greater than 30 volts DC. See the “Option Card Installation” section in the
Installation Chapter for information on how to access or remove the dis-
play module.
Содержание Thermox 2000 ATEX Series
Страница 18: ...1 8 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 24: ...2 6 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 68: ...3 44 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 128: ...5 28 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 149: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor...
Страница 151: ...Service and Parts 7 5 Figure 7 2 Cell replacement...
Страница 153: ...Service and Parts 7 7 Figure 7 3 Furnace thermocouple replacement...
Страница 157: ...Service and Parts 7 11 Figure 7 5 Furnace replacement...
Страница 159: ...Service and Parts 7 13 Figure 7 6 Combustibles detector replacement Hot Wire...
Страница 196: ...B 4 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 202: ...D 2 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...