Calibration | 5-1
CALIBRATION
Figure 5-1.
Calibrate Key menu
options.
The
Calibrate
key allows you to use the Series 2000 control unit to
calibrate your analyzer. The Series 2000 control unit provides features
that make calibrating your analyzer easy, and even includes a way for the
system to periodically calibrate itself.
Sensor alarms are disabled during calibrations and the recovery
time after a calibration. You can set whether analog outputs should
follow the calibration readings or hold at the last process reading
before the calibration.
NOTE
Initiate Cal
Initiate Verify
Cal/Verify Data
Cal Gas Value
Inject Cal Gas
Set Cal Timers
Calibrate Key
Sensitivity
Set K Factor
Содержание Thermox 2000 ATEX Series
Страница 18: ...1 8 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 24: ...2 6 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 68: ...3 44 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 128: ...5 28 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 149: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor...
Страница 151: ...Service and Parts 7 5 Figure 7 2 Cell replacement...
Страница 153: ...Service and Parts 7 7 Figure 7 3 Furnace thermocouple replacement...
Страница 157: ...Service and Parts 7 11 Figure 7 5 Furnace replacement...
Страница 159: ...Service and Parts 7 13 Figure 7 6 Combustibles detector replacement Hot Wire...
Страница 196: ...B 4 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 202: ...D 2 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...