Set Track/Hold Option
This menu option allows you to decide what signals to send to the
selected analog output port during a calibration or verification. You have
the option to either hold the last process reading or track calibration
readings. Separate decisions can be made for verify and calibrate opera-
tions, so you can hold the last process readings during verifications, but
track calibration gas readings during calibrations.
1. Select
Set Track/Hold
from the Analog Range key after selecting an
analog output port.
2. You can choose to either
hold
process readings or
track
calibration
readings on the analog output port. Separate choices can be made for
calibration and verify operations.
For calibration operations, select one of the following:
Hold during cal
or
Track during cal
After you select one of these options, your choice will be confirmed on
the display:
Out # 1 will TRACK
during Calibration
or
Out # 1 will HOLD
during Calibration
For verify operations, select one of the following:
Hold during verify
or
Track during verify
After you select one of these options, your choice will be confirmed on
the display:
Out # 1 will TRACK
during Verify
Содержание Thermox 2000 ATEX Series
Страница 18: ...1 8 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 24: ...2 6 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 68: ...3 44 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 128: ...5 28 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 149: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor...
Страница 151: ...Service and Parts 7 5 Figure 7 2 Cell replacement...
Страница 153: ...Service and Parts 7 7 Figure 7 3 Furnace thermocouple replacement...
Страница 157: ...Service and Parts 7 11 Figure 7 5 Furnace replacement...
Страница 159: ...Service and Parts 7 13 Figure 7 6 Combustibles detector replacement Hot Wire...
Страница 196: ...B 4 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 202: ...D 2 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...