Controller / User Interface | 4-25
NOTE
Alarm Key
Alarms 1 and 2 are hard wired for the watchdog timer and service
alarms and are therefore not software-selectable.
Oxygen and combustibles alarms are all disabled during a primary
or regular calibration. In addition, alarms are disabled during
start-up until the sensor reaches its proper operating temperature.
See the “Wiring” section in the Installation chapter for help on alarm
connections and how alarms are labeled on the control unit wiring card.
The Alarm key allows you to do the following:
y
Define whether Alarm 3 activates based on oxygen levels, the start of
a calibration, or the start of a verification (Select Function). All other
alarms are set to activate based on only a single condition; for Alarm
4 this condition is levels of oxygen, for Alarms 5 and 6 this condition
is high combustible levels (% or PPM scale).
Alarm Key
Alarm Set Points
Alarm Configure
Relay Configure
Select Function
Exception Log
Figure 4-7.
Alarm Key menu.
Содержание Thermox 2000 ATEX Series
Страница 18: ...1 8 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 24: ...2 6 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 68: ...3 44 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 128: ...5 28 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 149: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor...
Страница 151: ...Service and Parts 7 5 Figure 7 2 Cell replacement...
Страница 153: ...Service and Parts 7 7 Figure 7 3 Furnace thermocouple replacement...
Страница 157: ...Service and Parts 7 11 Figure 7 5 Furnace replacement...
Страница 159: ...Service and Parts 7 13 Figure 7 6 Combustibles detector replacement Hot Wire...
Страница 196: ...B 4 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 202: ...D 2 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...