Calibration | 5-7
Initiate Cal and Initiate Verify
The
Initiate Cal
menu option allows you to perform an automatic,
remote, or manual calibration. The
Initiate Verify
menu option allows
you to perform an automatic, remote, or manual verification. Because
these menu options function almost identically, we discuss them to-
gether. See the introduction to this chapter for more information on the
difference between a calibration and a verification.
Automatic calibrations and verifications can also be set to run at prede-
termined frequencies using the cycle times you define using the
Set Cal
Timers/Set Auto Timers
menu option (see later in this chapter for
details). Once you’ve set these cycle times, you don’t need to select the
Initiate Cal or Initiate Verify menu options for an automatic calibration to
occur.
The Auto Calibrate/Auto Verify and Remote Calibrate/Remote Verify
selections are only available if you have installed an Auto Calibrate Card
in your Series 2000 control unit and have a Remote Calibration Unit
(RCU). If the Series 2000 control unit detects that these options are not
installed, it will automatically prompt you with a manual calibration or
verification operation when you select either the
Initiate Cal
or
Initiate
Verify
menu options.
This section is divided into discussions on automatic, remote, and
manual calibrations or verifications. See the headings below to locate the
type of calibration or verification that applies to you, then follow those
instructions.
Содержание Thermox 2000 ATEX Series
Страница 18: ...1 8 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 24: ...2 6 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 68: ...3 44 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 128: ...5 28 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 149: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor...
Страница 151: ...Service and Parts 7 5 Figure 7 2 Cell replacement...
Страница 153: ...Service and Parts 7 7 Figure 7 3 Furnace thermocouple replacement...
Страница 157: ...Service and Parts 7 11 Figure 7 5 Furnace replacement...
Страница 159: ...Service and Parts 7 13 Figure 7 6 Combustibles detector replacement Hot Wire...
Страница 196: ...B 4 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 202: ...D 2 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...