6-18 | Thermox Series 2000 WDG-HPII / HPIIC - ATEX
LED
INDICATOR
STATUS FUNCTION
DS1
Oxygen cell furnace over-temperature shutdown
circuit status
GREEN
Furnace is operating normally
OFF
Over temperature shut-down has been activated
OFF
No 15 Volt DC power provided to sensor board from
control unit
DS2
Sensor enclosure internal temperature monitor
OFF
Oxygen-only system (no RTD and SW1 all positions
to “ON”)
Blinking GREEN
Sensor enclosure not at setpoint temperature
GREEN
Sensor enclosure at setpoint temperature
Blinking GREEN &
Blinking RED
Sensor enclosure over temperature
Blinking RED
Switch setup invalid
RED
Sensor temperature RTD circuit error (RTD open or
shorted)
YELLOW
Indeterminate
error
Sensor Board (80471SE) LEDs
Two diagnostic LED indicators are on the sensor board. DS1 is located near
the relay on the board and DS2 is located near the end of the numbered
terminal strip.
Содержание Thermox 2000 ATEX Series
Страница 18: ...1 8 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 24: ...2 6 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 68: ...3 44 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 128: ...5 28 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 149: ...Service and Parts 7 3 Figure 7 1 Wiring diagram for standard sensor...
Страница 151: ...Service and Parts 7 5 Figure 7 2 Cell replacement...
Страница 153: ...Service and Parts 7 7 Figure 7 3 Furnace thermocouple replacement...
Страница 157: ...Service and Parts 7 11 Figure 7 5 Furnace replacement...
Страница 159: ...Service and Parts 7 13 Figure 7 6 Combustibles detector replacement Hot Wire...
Страница 196: ...B 4 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...
Страница 202: ...D 2 Thermox Series 2000 WDG HPII HPIIC ATEX This page intentionally left blank...