7.3.3
Overload Light is On
CAUSE
SOLUTION
Unit is overloaded
Remove overload or check CL setting
7.3.4 Distorted
Output
CAUSE
SOLUTION
Power source is grossly overloaded.
Reduce load
The crest factor of the load exceeds 3:1.
Reduce load current peaks by reducing
load.
7.3.5
Unit Shuts Down after 1-2 Seconds
CAUSE
SOLUTION
Output open
Apply < 12 Ohm load or short for no load
Output grossly overloaded.
Remove overload.
-400 Input module failure
Have power module serviced
Operating load with too high inrush or start
up currents.
Consult factory for application advice.
7.3.6
No Output and No Lights on Front Panel
CAUSE
SOLUTION
Input circuit breaker switched off.
Switch the breaker on.
No input power.
Ensure 3 phase power is present at AC
input terminal block
-400 Input Power Supply failure
Have -400 input supply serviced.
7.3.7
No Output But Front Panel controller is active
CAUSE
SOLUTION
“OUTPUT ON” button is turned off.
Press OUTPUT ON so that “ON” LED is
lit.
Compliance voltage programmed down or
set to zero.
Program compliance voltage higher.
Current programmed down or to zero.
Turn current set point up.
Содержание 13500CS/3
Страница 5: ...User Manual Rev J CS Series California Instruments 5 SAFETY SYMBOLS ...
Страница 34: ...User Manual Rev J CS Series California Instruments 34 Figure 3 3 9000CS 2 Wiring diagram 3 Phase mode ...
Страница 35: ...User Manual Rev J CS Series California Instruments 35 Figure 3 4 13500CS 3 Wiring diagram 3 Phase mode ...
Страница 36: ...User Manual Rev J CS Series California Instruments 36 Figure 3 5 18000CS 4 Wiring diagram 3 Phase mode ...
Страница 56: ...User Manual Rev J CS Series California Instruments 56 The power source is now ready to be used ...
Страница 73: ...User Manual Rev J CS Series California Instruments 73 ...
Страница 111: ...User Manual Rev J CS Series CS Series 111 voltage rating 28 W Weight 22 Wiring AC input 30 ...