User Manual – Rev J
CS Series
California Instruments
106
Number
Message String
Cause
Remedy
42
"Voltage self test error,
output 3
No. 3/C amplifier in
Master source has no
output during Self-
test
Contact CI service department at
43
"Current self test error,
output 1
No. 1/A amplifier in
Aux. Source has no
output during Self-
test.
Contact CI service department at
44
"Current self test error,
output 2
No. 2/B amplifier in
Aux. Source has no
output during Self-
test.
Contact CI service department at
45
"Current self test error,
output 3
No. 3/C amplifier in
Aux. Source has no
output during Self-
test.
Contact CI service department at
216
" RS-232 receiver
framing error"
Communication
failure.
Check RS232 port settings and cable.
217
" RS-232 receiver parity
error"
Communication
failure.
Check RS232 port settings and cable.
218
" RS-232 receiver
overrun error"
Communication
failure.
Check RS232 port settings and cable.
402
"CAL password is
incorrect"
Calibration password
does not equal high
voltage range value.
Re-enter correct password.
403
"CAL not enabled"
No password entered
for calibration
Enter correct CAL password.
600
"Systems in mode:list
have different list
lengths"
Transient lists have
unequal lengths
Check list settings and correct to
same no of data points.
601
"Requested voltage and
waveform exceeds peak
voltage capability"
Wave shape selected
and RMS voltage
combine to exceed
peak voltage
capability.
Reduce RMS or crest factor of wave
shape.
602
"Requested voltage and
waveform exceeds
transformer volt-second
rating"
The selected wave
shape exceeds
output transformer
capability.
The volt-second product of he wave
form (magnitude and time in the +
and – half of wave form).
603
"Command only applies
to RS-232 interface"
Command not
relevant for GPIB
interface.
Do not use command.
604
"Trigger received before
requested number of pre-
trigger readings"
Data acquisition pre-
trigger buffer not filled
yet.
Hold off trigger or reduce pre-trigger
delay.
605
"Requested RMS current
too high for voltage
range"
Max RMS current is
function of voltage
range selected.
Reduce programmed RMS current
limit or select low voltage range.
606
"Waveform data not
defined"
No waveform name
specified
Specify waveform name before
sending waveform data.
607
"VOLT,VOLT:SLEW, and
FUNC:SHAPe modes
incompatible"
Conflict between
wave shape and
programmed slew
Reduce slew or change waveform
type.
Содержание 13500CS/3
Страница 5: ...User Manual Rev J CS Series California Instruments 5 SAFETY SYMBOLS ...
Страница 34: ...User Manual Rev J CS Series California Instruments 34 Figure 3 3 9000CS 2 Wiring diagram 3 Phase mode ...
Страница 35: ...User Manual Rev J CS Series California Instruments 35 Figure 3 4 13500CS 3 Wiring diagram 3 Phase mode ...
Страница 36: ...User Manual Rev J CS Series California Instruments 36 Figure 3 5 18000CS 4 Wiring diagram 3 Phase mode ...
Страница 56: ...User Manual Rev J CS Series California Instruments 56 The power source is now ready to be used ...
Страница 73: ...User Manual Rev J CS Series California Instruments 73 ...
Страница 111: ...User Manual Rev J CS Series CS Series 111 voltage rating 28 W Weight 22 Wiring AC input 30 ...