User Manual – Rev J
CS Series
California Instruments
105
Number
Message String
Cause
Remedy
-300
"Device specific error"
Hardware related
error
Check hardware for proper operation.
-311
"Memory error"
Waveform memory
checksum error.
May be the result of incomplete user-
defined waveform download. Check
interface and try downloading
waveform again. Successful
download may clear this error
condition.
Alternatively, use TRAC:DEL:ALL
command to clear waveform memory.
-314 "Save/recall
memory
lost"
User setup register
contents lost
Store setup in same register again.
-315 "Configuration
memory
lost"
Hardware
configuration settings
lost.
Contact CI service department at
to
obtain instructions on restoring
configuration data.
-330
"Self-test failed"
Internal error
Contact CI service department at
-350
"Queue overflow"
Message queue full.
Too many message. Read status
using SYST:ERR query until 0, "No
Error" is received indicating queue
empty.
-400
"Query error"
Unable to complete
query.
Check programming manual for
correct query format and parameters
-410
"Query INTERRUPTED"
Query issued but
response not read.
Check application program for correct
flow. Response must be read after
each query to avoid this error.
-420 "Query
UNTERMINATED"
Query incomplete.
Check for terminator after query
command.
-430 "Query
DEADLOCKED"
Query cannot be
completed
Check application program for
multiple queries
-440 "Query
UNTERMINATED"
Query incomplete.
Check for terminator after query
command.
0
"No error"
No errors in queue
2
" Non-volatile RAM
CONFIG section
checksum failed"
Controller failure
during Self-test.
Contact CI service department at
3
" Non-volatile RAM CAL
section checksum failed"
Controller failure
during Self-test.
Contact CI service department at
4
" Non-volatile RAM
WAVEFORM section
checksum failed"
Controller failure
during Self-test.
Contact CI service department at
10
"Ram self test
Controller failure
during Self-test.
Contact CI service department at
40
"Voltage self test error,
output 1
No. 1/A amplifier in
Master source has no
output during Self-
test.
Contact CI service department at
41
"Voltage self test error,
output 2
No. 2/B amplifier in
Master source has no
output during Self-
test
Contact CI service department at
Содержание 13500CS/3
Страница 5: ...User Manual Rev J CS Series California Instruments 5 SAFETY SYMBOLS ...
Страница 34: ...User Manual Rev J CS Series California Instruments 34 Figure 3 3 9000CS 2 Wiring diagram 3 Phase mode ...
Страница 35: ...User Manual Rev J CS Series California Instruments 35 Figure 3 4 13500CS 3 Wiring diagram 3 Phase mode ...
Страница 36: ...User Manual Rev J CS Series California Instruments 36 Figure 3 5 18000CS 4 Wiring diagram 3 Phase mode ...
Страница 56: ...User Manual Rev J CS Series California Instruments 56 The power source is now ready to be used ...
Страница 73: ...User Manual Rev J CS Series California Instruments 73 ...
Страница 111: ...User Manual Rev J CS Series CS Series 111 voltage rating 28 W Weight 22 Wiring AC input 30 ...