Number
Message String
Cause
Remedy
819
"Clock and sync must be
internal"
Operation not
possible with external
clock
Switch to internal sync. (Default)
820
"Input buffer full"
Too much data
received.
Break up data in smaller blocks.
821
"Amplifier unbalance"
Hardware error. An
amplifier has an
overload condition.
Check amplifier balance adjustment.
If error persists contact CI service at
.
822 "Waveform
harmonics
limit"
Harmonic contents of
user defined wave
shape is too high and
could damage
amplifier output
stage.
Reduce harmonic content or reduce
fundamental frequency programmed.
823
"Amplifier fault"
An amplifier failure.
Can be reported at
any time.
Determine which amplifier is at fault
with self-test or checking LED on
Relay Board. Replace amplifier.
824
“Auxiliary down”
One or more auxiliary
units is not powered
up or not working.
Turn on all auxiliary units.
825
“Over voltage prot trip”
Over voltage
detected on output
Check output voltage for correct RMS
value.
826
“Peak current prot trip”
Peak current limit
exceeded.
Peak current exceeded. Could be
caused by switching EUT on or off.
827
“Frequency error”
Frequency error
during self-test.
Correct frequency was not measured
during self-test. May be result of 801
error.
828
“Phase error”
Self test error phase
angle
Correct phase angle was not
measured during self-test. May be
result of 801 error.
829
“Dc component exceed
limit”
Too much DC content
in loaded ARB
waveform.
Check waveform programming.
Table 8-1: Error Messages
Содержание 13500CS/3
Страница 5: ...User Manual Rev J CS Series California Instruments 5 SAFETY SYMBOLS ...
Страница 34: ...User Manual Rev J CS Series California Instruments 34 Figure 3 3 9000CS 2 Wiring diagram 3 Phase mode ...
Страница 35: ...User Manual Rev J CS Series California Instruments 35 Figure 3 4 13500CS 3 Wiring diagram 3 Phase mode ...
Страница 36: ...User Manual Rev J CS Series California Instruments 36 Figure 3 5 18000CS 4 Wiring diagram 3 Phase mode ...
Страница 56: ...User Manual Rev J CS Series California Instruments 56 The power source is now ready to be used ...
Страница 73: ...User Manual Rev J CS Series California Instruments 73 ...
Страница 111: ...User Manual Rev J CS Series CS Series 111 voltage rating 28 W Weight 22 Wiring AC input 30 ...