User Manual – Rev J
CS Series
California Instruments
107
Number
Message String
Cause
Remedy
608
"Measurement
overrange"
Measurement data
out of range.
609
"Output buffer overrun"
Too much data in
output buffer.
Check receive mode on application
program. Program is not reading data
sent by AC source.
610
"Command cannot be
given with present
SYST:CONF setting"
Command conflicts
with available
hardware or firmware
option settings.
Check configuration for available
options and features.
801
"Output voltage fault"
- Output voltage
exceeds max
compliance voltage
set.
- Open circuit
Check load connection.
Load impedance may be too high for
programmed current level causing
compliance voltage to exceed set
limit.
802
"Current limit fault"
Maximum available
current exceeded.
Load exceeds current limit and unit is
in Constant Voltage (CV) mode of
operation. Reduce load or increase
CL setting
803
"Temperature fault"
Amplifier heat sink
temp. too high.
Reduce load. Ensure proper air flow
and exhaust clearance. Check fan(s)
for operation.
804
"External sync error"
Could not sync to
external sync signal.
External sync signal missing,
disconnected or out of range.
805
"Initial memory lost"
Initial settings could
not be recalled at
power-up.
Save power on settings again to
overwrite old content.
806
"Limit memory lost"
Hardware
configuration settings
could not be recalled
at power-up.
Contact CI service department at
to
obtain instructions on restoring
configuration data.
807
"System memory lost"
Memory corrupted
during power-up.
Recycle power.
808
"Calibration memory lost"
Calibration data lost
during power-up.
Contact CI service department at
to
obtain instructions on restoring
calibration data or recalibrate unit.
813
"Missing list parameter"
One or more transient
list parameters
missing.
Check programmed lists.
814
"Voltage peak error "
Peak voltage
exceeds internal bus
voltage
This error may occur when selecting
user defined wave shapes with higher
crest factors. Reduce programmed
RMS value.
815
"Slew time exceed dwell" Time needed to slew
to final value is less
than dwell time.
Check dwell times in transient list
settings. Increase dwell time or
change slew rate for affected
parameter.
816
"Illegal during transient"
Operation requested
not available while
transient is running.
Wait till transient execution is
completed or abort transient
execution first.
817
"Output relay must be
closed"
Transient
programmed with
output relay open.
Close relay before attempting
transient operation.
Содержание 13500CS/3
Страница 5: ...User Manual Rev J CS Series California Instruments 5 SAFETY SYMBOLS ...
Страница 34: ...User Manual Rev J CS Series California Instruments 34 Figure 3 3 9000CS 2 Wiring diagram 3 Phase mode ...
Страница 35: ...User Manual Rev J CS Series California Instruments 35 Figure 3 4 13500CS 3 Wiring diagram 3 Phase mode ...
Страница 36: ...User Manual Rev J CS Series California Instruments 36 Figure 3 5 18000CS 4 Wiring diagram 3 Phase mode ...
Страница 56: ...User Manual Rev J CS Series California Instruments 56 The power source is now ready to be used ...
Страница 73: ...User Manual Rev J CS Series California Instruments 73 ...
Страница 111: ...User Manual Rev J CS Series CS Series 111 voltage rating 28 W Weight 22 Wiring AC input 30 ...