Troubleshooting
Wrong lighting values or incorrect color in lighting menu
•
Place the lighting head over the ball bearing and verify if
each ring is complete. If not, replace the defective LED or
bank of LEDs.
•
Check if the right color is displayed for each ring. If not,
verify the wiring to the lighting controller.
•
Verify that unified_ligt_map.txt, SLC800LightingValues.xml,
SLC800Initialisation.xml and
LightingControllerIntensityTargets.xml are present and not
corrupted. Replace any missing or corrupted files (contact
Agilent Support).
•
Try to calibrate using a Fotowand gray card (see
).
Unable to auto-scale
•
When fiducial 1 is too close to the maximum or minimum
limits of the XY gantry inspection range, auto-scale cannot be
completed. Try auto-scaling on a different fiducial which is
not close to the stage limits.
Содержание U1840A Medalist sj5000
Страница 15: ...Safety and Regulatory Information 1 Agilent U1840A Medalist sj5000 Hardware Manual 1 9...
Страница 50: ...3 12 Agilent U1840A Medalist sj5000 Hardware Manual 3 Installation...
Страница 100: ...7 18 Agilent U1840A Medalist sj5000 Hardware Manual 7 Camera and Lighting Calibration...
Страница 134: ...10 18 Agilent U1840A Medalist sj5000 Hardware Manual 10 Setting Up Barcode Readers...
Страница 176: ...11 42 Agilent U1840A Medalist sj5000 Hardware Manual 11 Troubleshooting...
Страница 181: ...Maintenance 12 Agilent U1840A Medalist sj5000 Hardware Manual 12 5 Figure 12 2 Sensors on the conveyor...
Страница 190: ...A 4 Agilent U1840A Medalist sj5000 Hardware Manual A System Specifications...
Страница 192: ...B 2 Agilent U1840A Medalist sj5000 Hardware Manual B Reference Documents...