8-8
Agilent U1840A Medalist sj5000 Hardware Manual
8
Camera and Lighting Controller Testing
13
Keep clicking the
Level
button at the bottom right, until
Level 3 appears.
14
Perform the following checks:
NOTE
The camera must be in acquisition mode for the different lighting rings to
be triggered on and off.
Click this:
Check that feedback shows this:
90 Ring
01FF@F*
The red 90 degree ring in the lighting head should turn on.
70 Ring
02FF@0*
The red 70 degree ring should turn on and the 90 degree ring
turn off.
60 Ring
04FF@2*
The red 60 degree ring should turn on and the 70 degree ring
turn off.
45a Ring
40FF@2*
The red 45a degree ring should turn on and the 60 degree ring
turn off.
45b Ring
20FF@0*
The green 45b degree ring should turn on and the 45a degree
ring turn off.