Camera and Lighting Controller Testing
8
Agilent U1840A Medalist sj5000 Hardware Manual
8-7
The Terminal window should now show the following:
9
Keep clicking the
Level
button at the bottom right, until
Level 1 appears.
10
Click
SEG CURR
and check that the feedback is as follows:
*4SS0500@M*5SS0500@M*6SS0500@M*7SS0500@M*
11
Type in the following:
4SS469
5SS469
6SS469
7SS469
12
Click
Read
and check that the feedback is as follows:
‘RYSC@R1SC857,0,3757*@R2SC957,0,3757*@R3SC22257,0,37
57*@R4SC2457,469,3757*@R5SC2557,469,3757*@R6SC957,4
69,3757*@R7SC1757,469,3757@END*’
NOTE
The most important feedback result is when the READ button is clicked.
If there is no value in the middle of the each ring, as shown in the
example 4000*@R6SC700,0,4000, then this ring is not segmented. If the
feedback is 4000*@R6SC700,500,4000, then the ring is segmented.
Содержание U1840A Medalist sj5000
Страница 15: ...Safety and Regulatory Information 1 Agilent U1840A Medalist sj5000 Hardware Manual 1 9...
Страница 50: ...3 12 Agilent U1840A Medalist sj5000 Hardware Manual 3 Installation...
Страница 100: ...7 18 Agilent U1840A Medalist sj5000 Hardware Manual 7 Camera and Lighting Calibration...
Страница 134: ...10 18 Agilent U1840A Medalist sj5000 Hardware Manual 10 Setting Up Barcode Readers...
Страница 176: ...11 42 Agilent U1840A Medalist sj5000 Hardware Manual 11 Troubleshooting...
Страница 181: ...Maintenance 12 Agilent U1840A Medalist sj5000 Hardware Manual 12 5 Figure 12 2 Sensors on the conveyor...
Страница 190: ...A 4 Agilent U1840A Medalist sj5000 Hardware Manual A System Specifications...
Страница 192: ...B 2 Agilent U1840A Medalist sj5000 Hardware Manual B Reference Documents...