Agilent U1840A Medalist sj5000 Hardware Manual
7
Camera and Lighting Calibration
Auto Calibration
The auto calibration feature requires the use of a calibration
plate (optional, not included with the sj5000). Before starting
the auto calibration process, you must set the location of the
calibration plate.
1
Ensure that there are no boards in the sj5000.
2
From the
Tools
menu, click
Lighting Controller
under the
Camera
folder.
3
Click
Set Location
.
The system turns the LEDs on and displays the following
instruction:
4
Jog the camera as instructed and click
Set
.
5
Click
Yes
to confirm.
The system is now ready for lighting auto calibration.
6
Click
Auto Calibration
.
The camera automatically moves to the location of the
calibration plate and calibration will start.
7
You will be informed when the calibration is successfully
completed. Click
OK
.
The current outputs for each ring before and after calibration
are shown. These values area also written to the file
LightingController.log.
Содержание U1840A Medalist sj5000
Страница 15: ...Safety and Regulatory Information 1 Agilent U1840A Medalist sj5000 Hardware Manual 1 9...
Страница 50: ...3 12 Agilent U1840A Medalist sj5000 Hardware Manual 3 Installation...
Страница 100: ...7 18 Agilent U1840A Medalist sj5000 Hardware Manual 7 Camera and Lighting Calibration...
Страница 134: ...10 18 Agilent U1840A Medalist sj5000 Hardware Manual 10 Setting Up Barcode Readers...
Страница 176: ...11 42 Agilent U1840A Medalist sj5000 Hardware Manual 11 Troubleshooting...
Страница 181: ...Maintenance 12 Agilent U1840A Medalist sj5000 Hardware Manual 12 5 Figure 12 2 Sensors on the conveyor...
Страница 190: ...A 4 Agilent U1840A Medalist sj5000 Hardware Manual A System Specifications...
Страница 192: ...B 2 Agilent U1840A Medalist sj5000 Hardware Manual B Reference Documents...