Camera and Lighting Calibration
7
Agilent U1840A Medalist sj5000 Hardware Manual
7-9
Lighting Calibration
Lighting calibration sets the current output for each lighting
ring. It is performed over a gray card, which is used to calibrate
the illumination evenness. The gray card is manually loaded
into the sj5000. The system will then:
•
Turn each ring on.
•
Record the mean gray level intensity over the gray card.
•
Compare it to a target mean intensity.
•
Adjust the current output to bring the recorded mean as
close as possible to the target mean.
The new calculated means are referred to as Achieved Means
and are displayed in the Calibrated Means column in the
Lighting Controller window.
The procedures for lighting calibration are described in the
following sections:
•
: for manual calibration
•
: for automatic calibration
•
: at automatic calibration at specific
location
The calibration functions and lighting information are accessed
from the Lighting Controller window. For a summary of the
features, see
“The Lighting Controller Window”
NOTE
Lighting auto-calibration calculates all ring lighting modes available on
the sj5000.
Содержание U1840A Medalist sj5000
Страница 15: ...Safety and Regulatory Information 1 Agilent U1840A Medalist sj5000 Hardware Manual 1 9...
Страница 50: ...3 12 Agilent U1840A Medalist sj5000 Hardware Manual 3 Installation...
Страница 100: ...7 18 Agilent U1840A Medalist sj5000 Hardware Manual 7 Camera and Lighting Calibration...
Страница 134: ...10 18 Agilent U1840A Medalist sj5000 Hardware Manual 10 Setting Up Barcode Readers...
Страница 176: ...11 42 Agilent U1840A Medalist sj5000 Hardware Manual 11 Troubleshooting...
Страница 181: ...Maintenance 12 Agilent U1840A Medalist sj5000 Hardware Manual 12 5 Figure 12 2 Sensors on the conveyor...
Страница 190: ...A 4 Agilent U1840A Medalist sj5000 Hardware Manual A System Specifications...
Страница 192: ...B 2 Agilent U1840A Medalist sj5000 Hardware Manual B Reference Documents...