Control Module Screens of the Well-plate Samplers
8
1100 Series WS MWS Reference Manual
269
Maintenance Log
Use the
F4
-key (
Maintenance Log
) to look for maintenance activities
performed on the instrument. A list of possible events are listed in the scroll
screen. Select the activity you have performed and press
Done
to create an
entry in the logbook. For troubleshooting reasons this Logbook can be printed
or saved to a file on the PCMCIA card (pressing the
m
-key).
Firmware Update
Use the
Esc
-key to receive
Views
on the
F5
-key. Choose
System
from the
pull-down menu. Use the
F3
-key (
Records
) to select the WPS. Use the
F5
-key
(
FW Update
) to enter the Update section. If you want to update the resident
firmware (together with specific main firmware revisions), select the a file
from the PCMCIA card (RESnnnn.DLB) and press execute. If you want to
update the main firmware, press
F7
-key (
Transfer
) to turn the module into
the resident mode (LED on module should blink yellow).
Содержание G1367A
Страница 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Страница 4: ...4 1100 Series WS MWS Reference Manual...
Страница 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Страница 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Страница 168: ...168 1100 Series WS MWS Reference Manual 5 Parts and Materials Figure 23 Injection Valve Assembly 1 2 4 3 6 5...
Страница 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Страница 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Страница 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Страница 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Страница 284: ...284 1100 Series WS MWS Reference Manual Index...
Страница 285: ......