Troubleshooting and Test Functions
3
1100 Series WS MWS Reference Manual
87
Vessel Stuck to Needle
The vessel sticks to the needle when the needle moves up.
Possible causes
•
Closing mat to rigid/thick.
•
Bad X or Theta positioning and the needle sticks into the wall between two
holes.
•
Defective encoder on the needle carrier assembly.
Suggested Actions
✔
Check that the closing mat is not too thick.
✔
Exchange the needle carrier assembly.
✔
Change the sample transport assembly.
✔
Change the MTP main board.