130
1100 Series WS MWS Reference Manual
4
Repairing the Sampler
Exchanging Internal Parts
WA R N I N G
The following procedures require opening the main cover of the sampler. Always
ensure the sampler is disconnected from the line power when the main cover is
removed. The security lever at the power input socket prevents the autosampler
cover from being taken off when line power is still connected.
WA R N I N G
The power supply still uses some power, even if the power switch on the front panel
is turned off. To disconnect the sampler from line power, unplug the power cord.
WA R N I N G
When opening capillary or tube fittings solvents may leak out. Please observe
appropriate safety procedures (for example, goggles, safety gloves and protective
clothing) as described in the material handling and safety data sheet supplied by the
solvent vendor, especially when toxic or hazardous solvents are used.
N O T E
The electronics of the sampler will not allow operation when the top cover and the top foam
are removed. A safety light switch on the main board will inhibit the operation of the
sampler. Always operate the sampler with the top foam and top covers in place.
C A U T I O N
Internal components may be sensitive to electrostatic discharge (ESD). Always use an
ESD strap when handling internal components (see
Содержание G1367A
Страница 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Страница 4: ...4 1100 Series WS MWS Reference Manual...
Страница 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Страница 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Страница 168: ...168 1100 Series WS MWS Reference Manual 5 Parts and Materials Figure 23 Injection Valve Assembly 1 2 4 3 6 5...
Страница 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Страница 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Страница 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Страница 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Страница 284: ...284 1100 Series WS MWS Reference Manual Index...
Страница 285: ......