3
Troubleshooting and Test Functions
4. Rheodyne valve error
✔
Turn the system off and on twice
✔
Check the connectors on the SUD board
✔
Check the sampling unit connector on the main board
✔
Check the connector on the sampling unit
✔
Change the rheodyne valve (0101-0921)
✔
Change the sampling unit (G1367-60008)
✔
Change the main board (G1367-69500)
5. Needle lock error
✔
Check the connectors on the SUD board
✔
Check the sampling unit connector on the main board
✔
Check the connector on the sampling unit
✔
Change the sampling unit (G1367-60008)
✔
Change the main board (G1367-69500)