Repairing the Sampler
4
1100 Series WS MWS Reference Manual
133
Transport Assembly
Removing the sample transport
1
In the user interface start the maintenance mode and select the “Change
Needle/Seat” function.
2
Switch off the sampler at the main power switch. Remove the power cable.
3
Remove the top cover, top plate, and foam (see
4
Remove the loop capillary corrugated waste tube.
5
Disconnect the loop capillary from the analytical head (Use a 1/4 inch
wrench if you remove a SST capillary or a 4 mm wrench if you remove a
fused silica capillary).
6
Lift out the transport assembly. This may require a flat head screwdriver to
separate the sample transport assembly from the sampling unit.
Installing the sample transport
1
Slide the new transport assembly into the sampler.
2
Ensure the transport assembly is seated firmly.
3
Reconnect the loop capillary to the analytical head (Use a 1/4 inch wrench
if you reconnect a SST capillary or a 4 mm wrench if you reconnect a fused
silica capillary).
4
Re install the loop capillary corrugated waste tube
5
Install the top cover, top plate, and foam (see
6
Turn ON the sampler, close the front door, the system will reset
automatically.
7
In the user interface perform a “Sample Transport Self Alignment”.
When required
Defective transport assembly/Defective flex board or sensors.
Tools required
1/4 inch-5/16 inch wrench 8710-0510 (supplied in accessory kit)
4 mm open end wrench 8710-1534 (supplied in accessory kit)
Pozidrive screwdriver
Flat-head screwdriver (if interface board installed)
Parts required
G1367-60019 Sample Transport assembly
for the G1367A/68A
G1377-60009 Sample Transport assembly
for the G1377A/78A
Содержание G1367A
Страница 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Страница 4: ...4 1100 Series WS MWS Reference Manual...
Страница 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Страница 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Страница 168: ...168 1100 Series WS MWS Reference Manual 5 Parts and Materials Figure 23 Injection Valve Assembly 1 2 4 3 6 5...
Страница 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Страница 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Страница 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Страница 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Страница 284: ...284 1100 Series WS MWS Reference Manual Index...
Страница 285: ......