1100 Series WS MWS Reference Manual
3
Troubleshooting and Test Functions
Missing Vial or Missing Wash Vial
No vial was found in the position defined in the method or sequence.
When the needle carrier moves to a vial and the needle goes into the vial, the
position of the needle is monitored by an encoder behind the vial pusher. If no
vial is present, the encoder detects an error and the message “missing vial” is
generated.
Probable Causes
•
No vial in the position defined in the method or sequence.
•
Defective needle carrier assembly.
•
Defective transport assembly flex board.
•
Defective MTP board.
Suggested Actions
✔
Install the sample vial in the correct position, or edit the method or
sequence accordingly.
✔
Exchange the needle carrier assembly.
✔
Exchange the sample transport assembly.
✔
Exchange the MTP main board.
Содержание G1367A
Страница 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Страница 4: ...4 1100 Series WS MWS Reference Manual...
Страница 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Страница 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Страница 168: ...168 1100 Series WS MWS Reference Manual 5 Parts and Materials Figure 23 Injection Valve Assembly 1 2 4 3 6 5...
Страница 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Страница 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Страница 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Страница 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Страница 284: ...284 1100 Series WS MWS Reference Manual Index...
Страница 285: ......