1100 Series WS MWS Reference Manual
4
Repairing the Sampler
Installing the metering drive motor and belt
1
Feed the motor cable through to the SUD board. Connect the cable to
(labelled “Metering M”).
2
Install the motor with the four fixing screws.
3
Ensure the belt is seated correctly over the gear and motor shaft.
4
Install the sampling unit, (see
“Installing the sampling unit"
5
Install the sample transport, (see
“Installing the sample transport"
6
Install the top cover, top plate, and foam (see
7
Replace the plate tray in the tray base.
8
Turn ON the sampler, close the front door.
9
The system will reset automatically.
10
In the user interface perform a “Sample Transport Self Alignment”.