Repairing the Sampler
4
1100 Series WS MWS Reference Manual
153
Installing the power supply
1
Insert the coupler onto the switch of the new power supply.
2
Install the power supply in the module.
3
Connect the light pipe to the coupler.
4
Install the bottom foam. Ensure the cables are positioned properly.
5
Install the exhaust fan, (see
6
Install the MTP board, (see
“Installing the MTP main board"
7
Install the main fan, (see
8
Install the sampling unit, (see
“Installing the sampling unit"
9
Install the sample transport, (see
“Installing the sample transport"
10
Install the top cover, top plate, and foam (see
11
Replace the plate tray in the tray base.
12
Turn ON the sampler, close the front door.
13
The system will reset automatically.
14
In the user interface perform a “Sample Transport Self Alignment”.
Содержание G1367A
Страница 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Страница 4: ...4 1100 Series WS MWS Reference Manual...
Страница 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Страница 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Страница 168: ...168 1100 Series WS MWS Reference Manual 5 Parts and Materials Figure 23 Injection Valve Assembly 1 2 4 3 6 5...
Страница 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Страница 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Страница 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Страница 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Страница 284: ...284 1100 Series WS MWS Reference Manual Index...
Страница 285: ......