236
1100 Series WS MWS Reference Manual
7
Theory of Operation
One start bit and one stop bit are always used (not selectable).
Per default, the module will turn into 19200 baud, 8 data bit with no parity.
Forced Cold Start Settings
Switches 1 and 2 do not force storage of this set of parameters in non-volatile
memory. Returning switches 1 and 2 to other positions (other than being both
up) will allow for normal operation.
If you use the following switch settings and power the instrument up again, a
forced cold start has been completed.
To return to normal operation, set switches back to your GPIB or RS 232
configuration settings.
Table 74
Parity Settings
Switches
Parity
7
8
0
0
No Parity
1
0
Odd Parity
1
1
Even Parity
C A U T I O N
Forced cold start erases all methods and data stored in the non-volatile memory.
Exceptions are diagnose and repair log books which will not be erased.
Table 75
Forced Cold Start Settings
Mode Select
1
2
3
4
5
6
7
8
TEST/BOOT
1
1
0
0
0
0
0
1
Содержание G1367A
Страница 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Страница 4: ...4 1100 Series WS MWS Reference Manual...
Страница 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Страница 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Страница 168: ...168 1100 Series WS MWS Reference Manual 5 Parts and Materials Figure 23 Injection Valve Assembly 1 2 4 3 6 5...
Страница 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Страница 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Страница 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Страница 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Страница 284: ...284 1100 Series WS MWS Reference Manual Index...
Страница 285: ......