62
1100 Series WS MWS Reference Manual
3
Troubleshooting and Test Functions
Shutdown
An external instrument has generated a shut-down signal on the remote line.
The sampler continually monitors the remote input connectors for status
signals. A LOW signal input on pin 4 of the remote connector generates the
error message.
Probable Causes
•
Leak detected in another Agilent 1100 module with a CAN connection to
the system.
•
Error detected in an external instrument with a remote connection to the
system.
•
The degasser failed to generate sufficient vacuum for solvent degassing.
Suggested Actions
✔
Determine which Agilent 1100 module has the leak. Fix the leak before
restarting the well-plate sampler.
✔
Check external instruments for an error condition.
✔
Check the degasser for an error condition. Refer to the
Reference Manual
for the Agilent 1100 Series degasser.
Содержание G1367A
Страница 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Страница 4: ...4 1100 Series WS MWS Reference Manual...
Страница 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Страница 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Страница 168: ...168 1100 Series WS MWS Reference Manual 5 Parts and Materials Figure 23 Injection Valve Assembly 1 2 4 3 6 5...
Страница 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Страница 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Страница 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Страница 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Страница 284: ...284 1100 Series WS MWS Reference Manual Index...
Страница 285: ......