238
1100 Series WS MWS Reference Manual
7
Theory of Operation
The Main Power Supply Assembly
The main power supply comprises a closed assembly (no on site repair
possibility).
The power supply provides all DC voltages used in the module except for the
voltages supplied by the lamp power supply to the deuterium and tungsten
lamps in the detectors. The line voltage can vary in a range from 100 240 volts
AC ± 10 % and needs no manual setting.
To disconnect the instrument from line, unplug the power cord. The power
supply still uses some power, even if the power switch on the front panel is
turned off.
Figure 45
Main Power Supply (MPS) Blockdiagram
primary
circuits
+24 V
supply
+5 V
switching
linear
voltage
regulators
+12 V
supply
sense
MPS
Line
voltage
+36 V / 2.5 A (*
+24 V / 4.5 A (*
+5 V / 3 A
+15 V /.3 A
-15 V /.3 A
(*) total power consumption on +36 V and +24 V must not exceed 107 watts.
Содержание G1367A
Страница 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Страница 4: ...4 1100 Series WS MWS Reference Manual...
Страница 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Страница 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Страница 168: ...168 1100 Series WS MWS Reference Manual 5 Parts and Materials Figure 23 Injection Valve Assembly 1 2 4 3 6 5...
Страница 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Страница 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Страница 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Страница 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Страница 284: ...284 1100 Series WS MWS Reference Manual Index...
Страница 285: ......