26
1100 Series WS MWS Reference Manual
1
Installing the Sampler
Installing a Thermostatted Sampler
1
Place the thermostat on the bench.
2
Remove the front cover and route the condensation drain tube to the waste
bottle.
3
Install the LAN interface board in the sampler (if required), see
4
Remove the adhesive tape which covers the side and front doors.
5
Open the front door and remove the left side door.
6
Remove the transport protection foam.
7
Re-install the corrugated waste tube in the plastic port.
Preparation
Locate bench space
Provide power connections
Unpack the sampler and the thermostat
Parts required
Sampler and thermostat
Power cord, for the other cables see below and,
ChemStation and/or Control Module G1323B
WA R N I N G
Make sure that the condensation drain tube is always above the liquid level in the
vessel. If the tube is located in liquid the condensed water cannot flow out of the
tube and the outlet is blocked. Any further condensation will then remain in the
instrument. This may damage the instruments electronics.
Figure 6
Condensation leak outlet
Condensation drain tube
Waste bottle
Содержание G1367A
Страница 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Страница 4: ...4 1100 Series WS MWS Reference Manual...
Страница 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Страница 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Страница 168: ...168 1100 Series WS MWS Reference Manual 5 Parts and Materials Figure 23 Injection Valve Assembly 1 2 4 3 6 5...
Страница 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Страница 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Страница 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Страница 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Страница 284: ...284 1100 Series WS MWS Reference Manual Index...
Страница 285: ......