Troubleshooting and Test Functions
3
1100 Series WS MWS Reference Manual
63
Remote Timeout
A not-ready condition is still present on the remote input.
When an analysis is started, the system expects all not-ready conditions (e.g. a
not-ready condition during detector balance) to switch to run conditions
within one minute of starting the analysis. If a not-ready condition is still
present on the remote line after one minute the error message is generated.
Probable Causes
•
Not-ready condition in one of the instruments connected to the remote line.
•
Defective remote cable.
•
Defective components in the instrument showing the not-ready condition.
Suggested Actions
✔
Ensure the instrument showing the not-ready condition is installed
correctly, and is set up correctly for analysis.
✔
Exchange the remote cable.
✔
Check the instrument for defects (refer to the instrument’s reference
documentation).
Содержание G1367A
Страница 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Страница 4: ...4 1100 Series WS MWS Reference Manual...
Страница 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Страница 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Страница 168: ...168 1100 Series WS MWS Reference Manual 5 Parts and Materials Figure 23 Injection Valve Assembly 1 2 4 3 6 5...
Страница 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Страница 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Страница 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Страница 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Страница 284: ...284 1100 Series WS MWS Reference Manual Index...
Страница 285: ......