Troubleshooting and Test Functions
3
1100 Series WS MWS Reference Manual
71
Exhaust Fan Failed
The exhaust fan in the well-plate sampler has failed.
The hall sensor on the fan shaft is used by the WPS board to monitor the fan
speed. If the fan speed falls below a certain value the error message is
generated and the well-plate sampler shuts down.
Probable Causes
•
Fan cable disconnected.
•
Defective fan.
•
Defective MTP board.
Suggested Actions
✔
Ensure the fan is connected correctly.
✔
Exchange fan.
✔
Exchange the MTP board.