Theory of Operation
7
1100 Series WS MWS Reference Manual
233
Setting the 8-bit Configuration Switch
The 8-bit configuration switch is located next to the GPIB connector. Switch
settings provide configuration parameters for GPIB address, serial
communication protocol and instrument specific initialization procedures.
Switches 1 and 2 define which set of parameters (for example, for GPIB,
RS-232C, and so on) will be changed. Once the change has been completed, the
instrument must be powered up again in order to store the values in the
non-volatile memory.
In the non-volatile memory the parameters are kept, regardless of whether you
turn the instrument off and on again. They will be kept until the same set of
parameters is subsequently changed and power is reset. All other previously
stored configuration settings will still remain in the non-volatile memory.
Figure 44
8-bit Configuration Switch
Table 69
8-bit Configuration Switch
Mode Select
1
2
3
4
5
6
7
8
GPIB
0
0
GPIB Address
RS-232C
0
1
Baud rate
Data
Bits
Parity
Reserved
1
0
Reserved
TEST/BOOT
1
1
RSVD
SYS
RSVD
RSVD
FC
1
8
1
0
Содержание G1367A
Страница 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Страница 4: ...4 1100 Series WS MWS Reference Manual...
Страница 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Страница 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Страница 168: ...168 1100 Series WS MWS Reference Manual 5 Parts and Materials Figure 23 Injection Valve Assembly 1 2 4 3 6 5...
Страница 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Страница 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Страница 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Страница 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Страница 284: ...284 1100 Series WS MWS Reference Manual Index...
Страница 285: ......