Installing the Sampler
1
1100 Series WS MWS Reference Manual
39
Figure 13
Well-plate Dimensions (staggered)
Table 13
Well Plate Dimensions
Location Description
Definition
Limits
Rows
Number of rows on the plate
up to 16
Columns
Number of columns on the plate
up to 24
Volume
Volume (in µI) of a sample vessel
A
Row distance
Distance (in mm) between the
center of two rows
B
Column distance
Distance (in mm) between the
center of two columns
C
Plate length
X size (in mm) at the bottom of the
plate
127.75+/- 0.25 mm
(SBS Standard)
D
Plate width
Y size (in mm) at the bottom of the
plate
85.50+/-0.25 mm
(SBS Standard)
E
Plate height
Size (in mm) from the bottom to the
top of the plate
up to 47 mm
G
B
C
D
E
A
F
I
J
H
Содержание G1367A
Страница 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Страница 4: ...4 1100 Series WS MWS Reference Manual...
Страница 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Страница 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Страница 168: ...168 1100 Series WS MWS Reference Manual 5 Parts and Materials Figure 23 Injection Valve Assembly 1 2 4 3 6 5...
Страница 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Страница 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Страница 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Страница 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Страница 284: ...284 1100 Series WS MWS Reference Manual Index...
Страница 285: ......