86
1100 Series WS MWS Reference Manual
3
Troubleshooting and Test Functions
Vessel or Wash Vessel Error
The needle does not reach the target position in the vial or in the vessel of the
well-plate.
The sensor behind the vial pusher in the needle carrier assembly detects the
successful completion of the needle movement to the vessel. If the needle fails
to reach the end point, the sensor fails to recognize the needle movement and
the error message is generated.
Probable Causes
•
Bad vessel definition in the plate configuration.
•
Closing mats to rigid/thick.
•
Bad X or Theta positioning.
•
Defective encoder on the needle carrier assembly.
Suggested Actions
✔
Check the vessel definition in the plate configuration.
✔
Check that the closing mats is not too thick.
✔
Exchange the needle carrier assembly.
✔
Change the sample transport assembly.
✔
Change the MTP main board
Содержание G1367A
Страница 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Страница 4: ...4 1100 Series WS MWS Reference Manual...
Страница 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Страница 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Страница 168: ...168 1100 Series WS MWS Reference Manual 5 Parts and Materials Figure 23 Injection Valve Assembly 1 2 4 3 6 5...
Страница 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Страница 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Страница 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Страница 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Страница 284: ...284 1100 Series WS MWS Reference Manual Index...
Страница 285: ......