Troubleshooting and Test Functions
3
1100 Series WS MWS Reference Manual
103
8. Sample location error
✔
Check the plate configuration in the user interface
✔
Ensure the right vials and plates are used
✔
Perform an auto-alignment
✔
Check the connector from the needle carrier to the sample transport unit
✔
Check the sample transport connectors on the main board/sample
transport
✔
Check the connector on the bottom of the sample transport unit
✔
Change the needle (G1367-87200) and the seat (G1367-87101)
✔
Change the needle carrier assembly (G1367-60010)
✔
Change the sample transport unit (G1367-60019)
✔
Change the main board (G1367-69500)