128
1100 Series WS MWS Reference Manual
4
Repairing the Sampler
Peristaltic Pump
1
Remove the corrugated leak tubing.
2
Press the two clips on the front of the peristaltic pump.
3
Pull the pump forward off the motor shaft.
4
Disconnect the tubing leading to the wash port and the tubing coming from
the solvent bottle.
5
Connect the wash port tubing to the upper tubing of the new pump (use
sandpaper to get a good grip on the tubing).
6
Connect the tubing coming from the solvent bottle to the lower tubing of the
new pump.
7
Push the pump onto the motor shaft until the clips click into place.
8
Reinstall the corrugated leak tubing.
When required
Tubing blocked or broken.
Tools required
sand paper
Parts required
5065-4445 Peristaltic pump
N O T E
The peristaltic pump is a replaceable unit. The tubing inside the pump is not replaceable.
Содержание G1367A
Страница 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Страница 4: ...4 1100 Series WS MWS Reference Manual...
Страница 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Страница 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Страница 168: ...168 1100 Series WS MWS Reference Manual 5 Parts and Materials Figure 23 Injection Valve Assembly 1 2 4 3 6 5...
Страница 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Страница 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Страница 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Страница 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Страница 284: ...284 1100 Series WS MWS Reference Manual Index...
Страница 285: ......